ContaminationSensor CS 1000 Status Messages
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 89/112
BeWa CS1000 3764916 300 en-us 2012-08-29.doc 2012-08-29
LED Blink code /
Display /
Analog output /
Switch output
CS1000 Status To do
Error
no.
Red
4,1 mA / 2,05 V*
Open
Firmware error
Perform a reset. (To
do this, disconnect
the CS from the
voltage supply) or
contact HYDAC.
-1…-19
Red
4,1 mA / 2,05 V*
Open
Communic. error Check the wiring.
-20…-39
Red
4,1 mA / 2,05 V*
Open
System error
Perform a reset. (To
do this, disconnect
the CS from the
voltage supply) or
contact HYDAC.
-40…-69
Red
4,1 mA / 2,05 V*
Open
Error during
automatic
adjustment
Perform a reset. (To
do this, disconnect
the CS from the
voltage supply)
/ check the flow rate
or contact HYDAC.
-70
Red
4,1 mA / 2,05 V*
Open
Error measuring cell
LED
Perform a reset. (To
do this, disconnect
the CS from the
voltage supply)
/ check the flow rate
or contact HYDAC.
-100
* Is not valid for HDA 5500 output signal